3-19
VIBXPERT II 05.2012
Left:
Changing the measurement task
Right:
Task Manager
Setups for
measurement, sensor, evaluation and
RPM measurement
* Preconfigured measurement tasks/set-
ups are marked with a lock and cannot
be edited (except for transducer
& channel).
**Initially only the 'available transduc-
ers' appear in the transducer setup
(see P. 2-19). To display all transducers,
click on the bottom list entry '...more
transducers'
Open the 'Task Manager '
M = in Multimode only
Options
Use this option when you do not have an RPM sensor on the route and
need to perform measurements that require the RPM (e.g. roller bear-
ing condition, order-based spectra). If the measurement task already
has information on the RPM, this option is ignored.
Changing the measurement task (M)
• Highlightthemeasurementtypeintheselectionscreen.
• PresstheFkey.Alistappearswithallthemeasurementtasksset
up for this measurement type.
• Clickontherequiredmeasurementtask(seebelow).
Setting up a new measurement task (M)
VIBXPERT has a large selection of frequently used measurement
tasks* that can be extended by user-defined measurement tasks if
necessary. To configure a new measurement task, proceed as follows:
• Highlightthemeasurementtypeintheselectionscreen.
• PresstheMENUkeyandclickon'TaskManager'.TheTaskMan-
ager appears and can be used to set up the new measurement task
(seeP.3-2).
• Clickonthe'Measurementtask'pulldownmenu.
• PresstheMENUkeyandclickon'New'.Thetexteditorappears.
• Enteranameforthenewmeasurementtask.
• Changethemeasurementchannelifnecessary.
• Selectthesetupsformeasurement(seebelow),sensor**,evalua-
tionorRPMinthe'Setup'eldofthepulldownmenu.Ifthematch-
ing entry cannot be found, a new setup must be created (see next
section).
• ThenpresstheMENUkeyandclickon'OK'.