Remote Control
R&S
®
FPC
298User Manual 1178.4130.02 ─ 07
Prerequisites for this command
●
Select frequency scan ([SENSe:]FREQuency:MODE).
Parameters:
<StepSize> <numeric value>
Default unit: Hz
Example: //Define scan range with measurement every 10 MHz
FREQ:MODE SCAN
SCAN:STAR 30MHZ
SCAN:STOP 100MHZ
SCAN:STEP 10MHZ
Manual operation: See "Scan Step" on page 379
[SENSe:]SCAN:STOP <Frequency>
This command defines the stop frequency for a frequency scan.
Prerequisites for this command
●
Select frequency scan ([SENSe:]FREQuency:MODE).
Parameters:
<Frequency> <numeric value>
Default unit: Hz
Example: //Define scan range with measurement every 10 MHz
FREQ:MODE SCAN
SCAN:STAR 30MHZ
SCAN:STOP 100MHZ
SCAN:STEP 10MHZ
Manual operation: See "Stop Frequency" on page 379
[SENSe:]SWEep:SPACing <Scale>
This command selects the scale of the x-axis.
Parameters:
<Scale> LINear
Linear scaling of the x-axis.
LOGarithmic
Logarithmic scaling of the x-axis.
*RST: LINear
Example: //Select scaling of the x-axis
SWE:SPAC LOG
Manual operation: See "Linear Scale" on page 379
See "Log Scale" on page 380
Remote Commands of the Receiver Application