Cubis MSA User Manual 103
Weighing in Percent
t Define which percent calculation should be used.
Example: Weight value = 10 g, reference weight = 50 g
4GUKFWG: The result is the percentage share of the weight value of the
reference weight (Rest: Weight value/reference weight * 100%;in the example: 20%)
.QUU: The result is the percentage loss between the weight value and the refer-
ence weight (loss: weight value - reference weight / reference weight * 100%; in the
example: -80%)
4CVKQ&4 The result is the percentage ratio of the difference to the weight
value (Ratio 1: reference weight - weight value / weight value * 100%; in the
example: 400%)
4CVKQ14 The result is the percentage ratio of the reference weight to the
weight value (Ratio 2: reference weight / weight value * 100%; in the example:
500%)
t Define the number of decimal places to be used for the display of the percentage.
PQPG
t Define the accuracy to be used to apply the weight value for the reference
weight.
0QTOCNFKURNC[CEE
HQNFFGERNCEG
HQNFFGERNCEGU
y An overview of all settings is displayed.
t Check all settings and change as required.
y A message asks whether you would like to add another application.
t If required, select additional applications to be combined.
y Prompts for configuring weighing and printing appear.
t Check all settings and change as required.
t Enter a short name and a description for the new task.
t To save the new task, touch 5CXG.