46 Cubis MSA User Manual
System Settings (Menu)
Configuring Data Output (Print Parameters)
Here you can configure all settings for printing and data output.
Several of these settings depend on the interface configuration
(see chapter Interfaces).
h
Factory settings are marked with an *.
In order to receive detailed information on the individual setting options, touch
Info.
Interface to the print output: Options: COM port A, B, C (file/SD memory card),
D protocol: Options: Print/SBI/XBPI/Web service/SICS/Remote display.
This setting cannot be changed here, but is defined when configuring the inter-
faces.
h
Which parameters and settings are displayed here depends on the configured
protocol.
GLP standard print: When should the ISO/GLP output be possible? Options: *Off/
enabled for cal.-adjustment/always enabled.
Taring after printout: Automatic tare after printing? Options: *Off/on
Print event: What event should trigger printing?
Options: *PRINT key/Start of task/Initialization of application/Result of application/
Printing component/Evaluation of application/Exit task/Calibration - adjustment
Print key elements: Which elements should be printed?
h
For further information, please refer to the chapter Interfaces, in the section
Configuring a Printout.
Setting the Date and Time
t Enter the current date and time in the prescribed format.
Configuring Display and Control Unit Properties
The following properties of the display and control unit can be changed,
as necessary: .CPIWCIG, DCEMITQWPFEQNQTU, DTKIJVPGUU, CEQWUVKE
UKIPCN
.
These general user-independent properties are used when no user is logged in and
serve as a template when a new user is created.