94  Cubis MSA User Manual
Averaging
t  Enter the range of the unrest for the start of the measurement:
 HTQOOGCUWTGFQDLGEV
 This entry estimates the strength of unrest that is caused by the activity  
(e. g. animal movement). You can use the rough estimation or a more detailed  
estimation using percentages. 
 The weight value changes with each movement of the weighing pan. As soon 
as three consecutive measurement values are within the selected unrest range, 
the interval measurement begins.
Example: You are weighing animals with an average weight of approx. 300 g.
With the “5% from measured object” setting, the measurement starts when the 
movements are within a range of 15 g, i.e. ± 7.5 g. With the “50% from measured 
object” setting, the interval measurement would start when the weight value was 
within a range of 150 g, i.e. ± 75 g.
Activity
Digits
Measurement at 5% ±7,5 g
Start at minimum load 100 digits 
Time
Measurement at 50% ±75 g
400
350
300
250
200
150
100
50
0
 
t  Define the number of measurements to be used for the average value.
'PVGTHKZGFPWODGT: You must now enter a fixed number.
0WODGTECPDGEJCPIGF: You will define the number when you execute the 
task.