CLINITEK Status+ Analyzer Operator’s Guide 113
System Configuration
Note When you set the prompt for strip lot information, the system
also allows you to enter the cassette lot information before you run a
cassette test.
To set the prompt for lot information, perform the following steps:
1. On the Select Ready screen, select Instrument Set Up.
2. On the Choose Settings screen, select Instrument Settings.
3. On the Instrument Settings screen, select Urinalysis Test
Settings.
4. On the Urinalysis Test screen, select Next.
5. On the second Urinalysis Test screen, for Enter strip lot number
and expiration date before each test, select one of the following
options:
Enabled (default) prompts you to enter the strip or cassette
lot number and expiration date before each test.
Disabled does not display a prompt to enter strip or cassette
lot number and expiration date before each test.
6. Select Done 3 times to return to the Select Ready screen.
Editing the Reported Chemistries
By default, the system reports all chemistries in the test results: GLU,
BIL, KET, SG, BLO, pH, PRO, URO, NIT, LEU, ALB, and CRE. You can
include or exclude chemistries from the test results. The reported
chemistries you select apply to all Siemens strip types.
To edit the reported chemistries, perform the following steps:
1. On the Select Ready screen, select Instrument Set Up.
2. On the Choose Settings screen, select Instrument Settings.
3. On the Instrument Settings screen, select Urinalysis Test
Settings.
4. On the Urinalysis Test screen, select Edit reported chemistries.
5. On the Reported Chemistries 1 of 2 screen, change the reported
chemistries. By default, all chemistries are selected.
Select a chemistry to include it.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens CLINITEK Status+ and is the answer not in the manual?

Siemens CLINITEK Status+ Specifications

General IconGeneral
DisplayLCD
Sample TypeUrine
Measurement MethodReflectance photometry
TypeUrinalysis Analyzer
Test TypesUrine Chemistry
ConnectivityRS232 port
Weight5.0 kg
Power Supply100-240 V AC, 50-60 Hz
Data StorageUp to 200 patient results
Test ParametersLeukocytes, Nitrite, Urobilinogen, Protein, pH, Blood, Specific Gravity, Ketone, Bilirubin, Glucose
Test StripsMultistix and other compatible strips