76 CLINITEK Status+ Analyzer Operator’s Guide
Troubleshooting
The following table contains the issues that can occur when you
operate the analyzer.
Calling for Assistance
If your CLINITEK Status+ analyzer displays corrective actions for a
detected problem, carry out the instructions provided before you call
for assistance. If your actions do not correct the problem or the
instructions do not display, contact your local technical support
provider or distributor.
Technical Support
When you call for assistance about an error message, have the
following items ready. These items help the technical support
representative work on the issue as quickly as possible.
•Error number
Completed problem list (see Problem List‚ page 77)
Description Action
Display shows dashes Dashes in a field indicate where you
disabled a an option.
Dashes also display when you exclude
urinalysis tests for chemistries from the
test results.
You can write the information on the
blank lines in the test result printout, if
needed.
Test table movement is
irregular or slow
Heavy buildup of dried urine on the test
table.
Clean the test table and insert as
described in Section 4,
Maintenance, Weekly Cleaning of
the Test Table and Test Table Insert
page 57.
Low battery power.
Replace the batteries as described
in Section 4, Maintenance,
Changing the Batteries‚ page 63.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens CLINITEK Status+ and is the answer not in the manual?

Siemens CLINITEK Status+ Specifications

General IconGeneral
DisplayLCD
Sample TypeUrine
Measurement MethodReflectance photometry
TypeUrinalysis Analyzer
Test TypesUrine Chemistry
ConnectivityRS232 port
Weight5.0 kg
Power Supply100-240 V AC, 50-60 Hz
Data StorageUp to 200 patient results
Test ParametersLeukocytes, Nitrite, Urobilinogen, Protein, pH, Blood, Specific Gravity, Ketone, Bilirubin, Glucose
Test StripsMultistix and other compatible strips