CLINITEK Status+ Analyzer Operator’s Guide 55
Calibration & QC
Compare QC results to the QC manufacturer’s acceptable results list. If
the QC results are not acceptable, do not test the patient samples until
you solve the problem. Repeat QC tests until you have acceptable
results.
Quality Control Troubleshooting
If the control results fall outside the values stated in the product
instructions for use insert, try any of the following corrective actions:
Use a fresh strip out of a bottle or a fresh cassette out of a package
and repeat the QC test.
Use a fresh bottle of strips or fresh package of hCG cassettes to
repeat the quality control procedure.
Use a fresh control solution and repeat the quality control
procedure.
If you use lyophilized controls, prepare the control solution using a
fresh bottle of control product.
For more troubleshooting information, see Section 5, Troubleshooting,
or contact your local technical support providerfor assistance.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens CLINITEK Status+ and is the answer not in the manual?

Siemens CLINITEK Status+ Specifications

General IconGeneral
DisplayLCD
Sample TypeUrine
Measurement MethodReflectance photometry
TypeUrinalysis Analyzer
Test TypesUrine Chemistry
ConnectivityRS232 port
Weight5.0 kg
Power Supply100-240 V AC, 50-60 Hz
Data StorageUp to 200 patient results
Test ParametersLeukocytes, Nitrite, Urobilinogen, Protein, pH, Blood, Specific Gravity, Ketone, Bilirubin, Glucose
Test StripsMultistix and other compatible strips