74 CLINITEK Status+ Analyzer Operator’s Guide
Troubleshooting
Troubleshooting the Analyzer Operation
The following table contains the analyzer operation icons that can
display near the title bar on the Select Ready screen when an
operation issue occurs.
Icon Description Action
Low Battery Power Displays on the Select Ready
screen, indicating that the battery
power level is low. An advisory
message also displays when the
battery power level is low. The
power level decreases while the
testing continues.
If the battery level falls too low to
power the analyzer, you cannot run
a test until you replace the batteries.
Replace the batteries. For
instructions, see Section 4,
Maintenance, Changing the
Batteries‚ page 63.
No Printer Paper Displays on the Print Help button on
the Select Ready screen, indicating
that the printer is out of paper or a
label roll. An advisory message also
displays.
Replace the empty paper or
label roll with a new one, as
instructed in Section 1,
Introduction, Loading the
Printer Paper‚ page 14.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens CLINITEK Status+ and is the answer not in the manual?

Siemens CLINITEK Status+ Specifications

General IconGeneral
DisplayLCD
Sample TypeUrine
Measurement MethodReflectance photometry
TypeUrinalysis Analyzer
Test TypesUrine Chemistry
ConnectivityRS232 port
Weight5.0 kg
Power Supply100-240 V AC, 50-60 Hz
Data StorageUp to 200 patient results
Test ParametersLeukocytes, Nitrite, Urobilinogen, Protein, pH, Blood, Specific Gravity, Ketone, Bilirubin, Glucose
Test StripsMultistix and other compatible strips

Related product manuals