120 CLINITEK Status+ Analyzer Operator’s Guide
System Configuration
•Operator name
•Patient name
• Instrument serial number
•Patient ID
• Urine color
• Urine clarity
• Up to 2 header lines of customized alphanumeric text
You also can specify the following printer options:
• Enable or disable an external printer
• Enable or disable the internal printer
• Print sample interference notes
To customize the printout, perform the following steps:
1. On the Select Ready screen, select Instrument Set Up.
2. On the Choose Settings screen, select Instrument Settings.
3. On the Instrument Settings screen, select Printer Settings.
4. On the Printer Settings Included in print-out, 1 of 4 screen,
select or clear any of the following options:
•Operator Name
•Patient Name
• Serial Number
•Patient ID
5. Select Next. On the Printer Settings Included in print-out, 2 of 4
screen, select or clear any of the following options:
• Color
•Clarity
• Custom Information (Up to 2 header lines of customized
alphanumeric text)
6. Select Next.
7. On the Printer Settings Included in print-out, Set Up Custom
Header, 3 of 4 screen, perform the following steps:
•Select Enabled to include the custom header in the printout.