78 CLINITEK Status+ Analyzer Operator’s Guide
Troubleshooting
5. Does the display show the Select Ready screen or
the Results screen as expected?
6. Does the test table move into and out of the
analyzer?
7. Does a quality control solution give the expected
result?
8. Is the name of the Siemens Healthcare Diagnostics
urinalysis strip or Clinitest immunoassay cassette
shown on the display the same as the product being
used?
9. Does the display or printout show the correct test
names and expected results?
10. Is the white calibration bar on the test table dirty,
scratched, or damaged?
11. Additional problem observations, please describe:
_____________________________________________
_____________________________________________
_____________________________________________
_____________________________________________
_____________________________________________
ClinitekStatus+ Analyzer Problem List

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens CLINITEK Status+ and is the answer not in the manual?

Siemens CLINITEK Status+ Specifications

General IconGeneral
DisplayLCD
Sample TypeUrine
Measurement MethodReflectance photometry
TypeUrinalysis Analyzer
Test TypesUrine Chemistry
ConnectivityRS232 port
Weight5.0 kg
Power Supply100-240 V AC, 50-60 Hz
Data StorageUp to 200 patient results
Test ParametersLeukocytes, Nitrite, Urobilinogen, Protein, pH, Blood, Specific Gravity, Ketone, Bilirubin, Glucose
Test StripsMultistix and other compatible strips

Related product manuals