SIREMOBIL Iso-C SPR2-230.061.01 Page 10 of 10 Siemens AG
System Manual Rev. 13 12.04 CS PS 24 Medical Solutions
7 - 10 Brakes / Movements
• After the counterbalance weights have been reinstalled on both sides, move the C-arm
orbitally from one end to the other several times.
• In the case of SIREMOBIL Iso-C 3D, additionally perform a scan.
• If the belt tension is still insufficient, repeat the procedure and place the belt into the next
tooth to shorten it further.