1 - 1
Siemens AG SPR2-230.061.01 Page 1 of 18 SIREMOBIL Iso-C
Medical Solutions Rev. 13 12.04 CS PS 24 System Manual
Prerequisites 1
Required documents 1
• SIREMOBIL Iso-C circuit diagram
• Circuit diagram for monitor X2080
• SIREMOBIL Iso-C logbook
Required tools, measurement and auxiliary devices 1
• For the laser targeting device where used: Adjustment Instructions / Laser targeting
device RXR2-130.032.01...
For SIREMOBIL Iso-C 3D additional
• SIREMOBIL Iso-C 3D Startup Instructions SPR2-230.815.01...
• Image Quality Quick Test SPR2-230.037.01 Revision 05 or higher
• DMM e.g. Fluke 8060A 97 02 101
• Oscilloscope > 50 MHz e.g. TFluke Combi Scope PM 3390A 99 00 861
• Dose measurement device e.g. PTW-DIADOS 97 17 612
• Safety tester e.g. BENDER Unimet 1100 ST 51 38 727
• Service PC with WINDOWS 3.1., 3.11 or 95 operating system
• Service software (refer to SIREMOBIL Iso-C logbook)
• Connection cable, Service-PC - SIREMOBIL Iso-C e.g. 99 00 440
• 1 set of resolution tests e.g. 28 71 820
• 1 set of radiation filters e.g. 97 98 596
• Center cross e.g. 96 60 051
• Heat conducting paste WPS e.g. 20 48 650
• Special purpose oil (Optimol Optipit) 55 07 525
• Special purpose oil (Slic Pac PTFE) 55 07 517
• Sealing compound 20 49 716
• Standard service equipment*
• 1 set Allen keys*
• 200 N spring scale e.g. 44 15 113
• Torque wrench 20 Nm to 100
Nm
e.g. 80 86 159
• Torque wrench 3 Nm to 20 Nm e.g. 34 24 553
• A 30 mm socket key (is required for adjusting the brake)*
For SIREMOBIL Iso-C 3D additional
• Calibration Phantom Iso-C 3D (without navigation interface)
or Calibration Phantom Iso-C 3D (with navigation interface)
71 39 947
75 51 620
• Luminance meter e.g. SMfit light field lum. meter 88 81 281