SIREMOBIL Iso-C SPR2-230.061.01 Page 12 of 48 Siemens AG
System Manual Rev. 13 12.04 CS PS 24 Medical Solutions
6 - 12 Replacing boards / replacing components
Check the display of the blades on the monitor screen and adjust, if necessary.
Check collimation for direct exposure if there is a cassette holder.
Reinstall the rubber stop.
Ensure that the area/dose product measurement device is functioning properly
(if present).
Check the setting of the laser targeting device and, if necessary, adjust it according to
the Adjustment Instructions for the Laser Targeting Device, RXR2-130.032.01.
If the "MODULARIS" system is present, perform the isocenter check.
Replacing the I.I. 6
Checking the temperature indicator 6
Prior to installing the I.I., check the temperature indicator.
If the inner square field of the indicator is white, the I.I. did not exceed the temperature
range. Remove the temperature indicator.
If the indicator is discolored (inner field black), proceed according to the IQ document
RXD-000.038.01.
Electrical voltage!
See Chapter 1, Safety Information.
Switch the SIREMOBIL Iso-C off and wait at least > 3 minutes until
the high voltage in the system and in the I.I. mini voltage supply
dissipates.
WARNING