SIREMOBIL Iso-C SPR2-230.061.01 Page 12 of 48 Siemens AG
System Manual Rev. 13 12.04 CS PS 24 Medical Solutions
6 - 12 Replacing boards / replacing components
• Check the display of the blades on the monitor screen and adjust, if necessary.
• Check collimation for direct exposure if there is a cassette holder.
• Reinstall the rubber stop.
• Ensure that the area/dose product measurement device is functioning properly
(if present).
• Check the setting of the laser targeting device and, if necessary, adjust it according to
the Adjustment Instructions for the Laser Targeting Device, RXR2-130.032.01.
• If the "MODULARIS" system is present, perform the isocenter check.
Replacing the I.I. 6
Checking the temperature indicator 6
• Prior to installing the I.I., check the temperature indicator.
• If the inner square field of the indicator is white, the I.I. did not exceed the temperature
range. Remove the temperature indicator.
• If the indicator is discolored (inner field black), proceed according to the IQ document
RXD-000.038.01.
Electrical voltage!
See Chapter 1, Safety Information.
Switch the SIREMOBIL Iso-C off and wait at least > 3 minutes until
the high voltage in the system and in the I.I. mini voltage supply
dissipates.
WARNING