Adjustments / Programs 5 - 21
Siemens AG SPR2-230.061.01 Page 21 of 30 SIREMOBIL Iso-C
Medical Solutions Rev. 13 12.04 CS PS 24 System Manual
Calibrating the area dose product measurement device 5
CONSTANT parameter (Qt * Kg) = T
w
5
• Read and record the T
w
value shown on the electronics unit of the area dose product
measurement device.
Programming 5
• Connect the service PC to the serial interface of the SIREMOBIL Iso-C.
• Start the service program.
• Select the Adjustment, Parameter menu.
• In the Combo box "Parameter Groups:", select "Diamentor" .
• Click on "Get from Unit".
• The Parameter "1. Diamentor Existing" must be on YES .
• Select parameter "2. Constant (Qt * Kg)" .
• In "Value Actual:"field, enter the parameterT
w
.
• Click on "Set Value".
• Click on "Put to Unit" .
• Click on "Save to File".
Fig. 6
This adjustment is used to calibrate the area dose product mea-
surement device of the SIREMOBIL Iso-C display.
You must exit the service program with "Logoff" and "Quit" to
store the modified parameters to disk as a new file.
Potentiometer
NOTE
NOTE