SIREMOBIL Iso-C SPR2-230.061.01 Page 16 of 48 Siemens AG
System Manual Rev. 13 12.04 CS PS 24 Medical Solutions
6 - 16 Replacing boards / replacing components
Concluding work 6
• Reinstall the I.I. on the C-arm flange. Refer to chapter 1, I.I. service position.
• Check the collimator display on the monitor and, if necessary, adjust it.
• Perform the I.Q. Quick test.
• Check the setting of the laser light localizer, if present. If necessary, adjust it.
• If the "MODULARIS" system is present, perform the isocenter check.
• Execute the country-specific acceptances (§16 Partial acceptance test, DHHS,...).