SIREMOBIL Iso-C SPR2-230.061.01 Page 16 of 48 Siemens AG
System Manual Rev. 13 12.04 CS PS 24 Medical Solutions
6 - 16 Replacing boards / replacing components
Concluding work 6
Reinstall the I.I. on the C-arm flange. Refer to chapter 1, I.I. service position.
Check the collimator display on the monitor and, if necessary, adjust it.
Perform the I.Q. Quick test.
Check the setting of the laser light localizer, if present. If necessary, adjust it.
If the "MODULARIS" system is present, perform the isocenter check.
Execute the country-specific acceptances (§16 Partial acceptance test, DHHS,...).