Replacing boards / replacing components 6 - 23
Siemens AG SPR2-230.061.01 Page 23 of 48 SIREMOBIL Iso-C
Medical Solutions Rev. 13 12.04 CS PS 24 System Manual
Place an 8 mm washer over the three markings. All three markings must be located with-
in the 8 mm diameter of the washer. An 8 mm washer has a somewhat larger inner diam-
eter of 8.4 mm, therefore the markings must clearly lie within the inner diameter and
should not touch the edge of the washer. If necessary, repeat the adjustment, until the
three markings fall within the 8 mm diameter.
Secure the camera optics.
Retighten the eccentric screws of the camera optics.
Checks and adjustments 6
Check the camera rotation and, if necessary, adjust it.
Check the setting of the radiation iris and, if necessary, readjust it.
Check the display of the blades on the monitor and adjust, if necessary.
Check the dose rate again and, if necessary, adjust it.
Perform the IQ quick test.
If the "MODULARIS" is present, perform the isocenter check.