Replacing boards / replacing components 6 - 27
Siemens AG SPR2-230.061.01 Page 27 of 48 SIREMOBIL Iso-C
Medical Solutions Rev. 13 12.04 CS PS 24 System Manual
• Configure the following parameters in User setup for MEMOSKOP:
Monitor zoom Configure No.
Monitor split Configure Yes.
Monitor text Configure Yes.
Maximum image number Configure 100.
Disable doserate selection
in user setup
Set to "No" when equipped with an area dose product mea-
surement device. The menu "Storage of Dose on disk" and
the menu "Dose on Hardcopy output" in User Setup are visi-
ble. Otherwise set to "Yes".
Unit serial number Program the serial number of the SIREMOBIL.
Do not enter the leading zero.
Product name Program "SIREMOBIL Iso-C".
Default program change-
able
Default: Yes.
Depending on country specific regulations configure Yes or
No.
Create external disk Configure No.
Default LUT Configure 2. Program a different LUT curve only upon cus-
tomers request.
Default Edge Configure 1. Program a different Edge enhancement value
only upon customers request.
Software backward com-
patible
Configure No.
Operating mode Para-
meters
Change the programmed parameters only upon customer
request.
VCR output Configure to monitor A.
Hardcopy Release For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors: Configure to monitor B.
Text monitor For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors: configure to both monitors.
Split screen Configure horizontally. Configure vertically only upon cus-
tomer request.
Display of time Configure Yes.
Set Date Time Set the current date and time.
Hospital Name Program the hospital name.
Storage of Dose on disk If area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
Dose on Hardcopy output If area dose product meter is present, (Diamentor) con-
figure yes. Otherwise configure No.
• Perform IQ quick test.