SIREMOBIL Iso-C SPR2-230.061.01 Page 30 of 48 Siemens AG
System Manual Rev. 13 12.04 CS PS 24 Medical Solutions
6 - 30 Replacing boards / replacing components
Configure the following parameters in User setup for MEMOSKOP:
Disable doserate selection
in user setup
Set to "No" when equipped with an area dose product mea-
surement device. The menu "Storage of Dose on disk" and
the menu "Dose on Hardcopy output" in User Setup are visi-
ble. Otherwise set to "Yes".
Unit serial number Program the serial number of the SIREMOBIL.
Do not enter the leading zero.
Default LUT Configure 2. Only upon customers request program a
different LUT curve.
Default Edge Configure 1. Only upon customers request program a
different Edge enhancement value.
Software backward com-
patible
Configure No.
Product name Program "SIREMOBIL Iso-C".
Default program change-
able
Default: Yes.
Depending on country specific regulations configure Yes or
No.
Create external disk If the option "3D Reconstruction" or the option
"DICOM Bridge" is installed or a MOD drive is present, con-
figure Yes. Otherwise configure No.
Enable Bridge If the option "3D Reconstruction" or the option
"DICOM Bridge" is installed or a MOD drive is present, con-
figure Yes. Otherwise configure No.
Delay 3D-PC/ DICOM
Bridge
Only available of if option "3D Reconstruction" is installed.
Configure 120.
Interface to 3D-PC Only available if option "3D Reconstruction" is installed.
Configure yes.
Operating mode Para-
meters
Change the programmed parameters only upon customer
request.
VCR output Configure to monitor A.
Hardcopy Release For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors: Configure to monitor B.
Text monitor For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors:
Configure to both monitors.
Split screen Configure horizontally. Configure vertically only upon cus-
tomer request.
Display of time Configure Yes.
Set Date Time Set the current date and time.
Hospital Name Program the hospital name.
Storage of Dose on disk If area dose product meter is present, (Diamentor)
configure Yes. Otherwise configure No.