Replacing boards / replacing components 6 - 33
Siemens AG SPR2-230.061.01 Page 33 of 48 SIREMOBIL Iso-C
Medical Solutions Rev. 13 12.04 CS PS 24 System Manual
Configure the following parameters in User setup for MEMOSKOP:
Perform the IQ quick test.
Roadmap K-Factor, phase B Program "MD 2".
Program a different value only upon customer request.
Roadmap K-Factor, phase C Program "MD 2".
Program a different value only upon customer request.
Roadmap Time of phase A Program "4,5".
Program a different value only upon customer request.
Operating mode Parameters Change the programmed parameters only upon customer
request.
VCR output Configure to monitor A.
Hardcopy Release For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors: Configure to monitor B.
Text monitor For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors:
configure to both monitors.
Split screen Configure horizontally.
Configure vertically only upon customer request.
Display of time Configure Yes.
Set Date Time Set the current date and time.
Hospital Name Program the hospital name.
Storage of Dose on disk If area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
Dose on Hardcopy output If area dose product meter is present, (Diamentor) con-
figure yes. Otherwise configure No.
With present System MODULARIS the parameters for the Litho-
program have to be set again according the Service instructions
LITHOSTAR MODULARIS.
NOTE