SSI-6000/SSI-5800/SSI-5500/SSI-5500BW
Digital Color Doppler Ultrasound System
IEC 60601-2-37: 2001 + A1:2004 + A2:2005, Medical Electrical Equip-
ment Part 2-37: Particular Requirements For The Safety Of Ultrasonic
Medical Diagnostic And Monitoring Equipment
The equipment conforms with the following EMC/EMI standards: IEC60601-
1-2: 2001+A1: 2004, Class A (CE)
Degrees of protection against harmful liquid:
For the main sys-
tem:
IPX0
For the trans-
ducer:
IPX7, from the acoustic window to the junction line
(Figure 13.1);
and IPX1 for other parts that may contact with the
patient, excluding the transducer connector.
For maximum safety, adhere to these guidelines:
Proper grounding of the system is critical to avoid electrical shock. For pro-
tection, ground the system with a three-conductor cable, and plug the sys-
tem into a hospital-grade receptacle or outlet.
Do not remove or entangle the grounding wire.
Do not remove the protective covers of the system. These covers protect
users from hazardous voltages. Only authorized service technicians can
make replacements of the parts inside the system.
Never use this system while inflammable gas is present!
Devices must be powered from a receptacle marked “hospital grade” be-
fore being connected to the system directly. In case “hospital grade” re-
ceptacles are not available, use isolation transformers instead.
1.9 Transducer Maintenance
The transducers/probes provided with the system are durable and have reliable
performance. These precision instruments should be inspected daily and handled
with care. Please observe the following PRECAUTIONS:
Do not drop the transducer on the hard surface. This can damage the trans-
ducer elements and compromise the electrical safety of the transducer.
Avoid kinking or pinching the transducer cable.
Use only approved ultrasound coupling gels.
Follow the instructions for cleaning and disinfecting the probes.
Disinfecting Surface Transducers
Disconnect the transducer from the system.
Wipe all surfaces with isopropyl alcohol solution and air dry.
Clean all surfaces of the probe and cable with soft cloth.
P/N: 4701-0061-01B
1-11