6 Operation
38
Operator Manual
V-PRO
®
maX 2
10085896_H
DANGER
PERSONAL INJURY, CONTAMINATED LOAD AND/OR EQUIPMENT DAMAGE HAZARD
Processing a load containing a stainless-steel lumened instrument through the Non
Lumen or Fast Non Lumen (if equipped) Cycles results in an ineffective Sterilization
Cycle. Do not process stainless-steel lumened devices in a Non Lumen or Fast Non
Lumen Cycle.
Failure to thoroughly clean, rinse and dry articles to be Sterilized could result in an
ineffective Sterilization Cycle.
WARNING
ABORTED CYCLE HAZARD
A leak test is recommended to be performed monthly on the Sterilization Unit to ensure
chamber leak rate is within specification. The leak rate must be less than
0.3 Torr/minute (0.4 mbar/minute or 0.04 kPa/minute). If not, an abort results, door
remains locked and Sterilization Unit service is required.
WARNING
PERSONAL INJURY, CONTAMINATED LOAD AND/OR EQUIPMENT DAMAGE HAZARD
When closing the chamber door, keep hands and arms out of the door opening and
ensure opening is clear of any obstructions.
The Low Temperature Sterilization System is specifically designed to only process goods using the cycles presented in
Hydrogen Peroxide Sterilization. If there is any doubt about a specific material or product, contact the product
manufacturer for recommended technique. Proceed as follows:
NOTE: For a VHP Sterilization Cycle, note the following:
If Lumen Cycle, Non Lumen Cycle, Fast Non Lumen Cycle, or Flexible Cycle touch pad is pressed while the chamber door is
open, a display screen appears directing the operator to close the door before continuing cycle operation.
Control through the display panel alerts the user to replace an invalid, expired or empty Sterilant Cup prior to starting a
cycle. Refer to Sterilant Cup Installation and Removal, for more information.
If an alarm occurs during cycle operation, refer to Troubleshooting, for instructions on correcting the alarm condition.
1. Before running VHP Cycle, see Before Operating Sterilization Unit and Load Sterilization Unit. Ensure
Sterilization Unit is properly set up for operation.
2. From SELECT CYCLES (or Ready) screen (see Figure 6-4), operator may perform one of the following functions:
NOTE: For Control Display touch panel screen touch pad definitions, refer to Symbols.
a. Press Standby touch pad and IN STANDBY screen displays.