EasyManua.ls Logo

Steris V-PRO maX 2 - Page 57

Steris V-PRO maX 2
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Operation
46
Operator Manual
V-PRO
®
maX 2
10085896_H
NOTE: For an Aborted Cycle, note the following:
a. If the cycle is aborted before the Sterilization phase, the control system safely relieves chamber vacuum before
allowing opening of the chamber door. If the cycle is aborted during or after Sterilization phase, a complete
Aeration phase is performed.
b. Press lower right touch pad to silence alarm.
c. Pressing Cycle Status touch pad (lower left) displays Alarm Status Screen.
d. Pressing Back Arrow touch pad (lower left) returns CYCLE (or Ready) screen to control display.
3. When Abort phase is complete, ABORT Complete screen is displayed.
NOTE: STERIS recommends (in accordance with ANSI/AAMI ST58, 2013) wearing chemical-resistant gloves
when removing items from the Sterilization Unit after a cycle has been aborted.
4. Once chamber door is opened, display returns to CYCLE (or Ready) screen and Sterilization Unit may be unloaded.
IMPORTANT: The load must be repackaged and reprocessed after the cycle is aborted.
NOTE: ALL Cycles abort to AERATE phase on a Power Failure.
Pop-Up Screens
During Start-up and Cycle Operation, the Sterilizer control informs the operator of various situations that could impact
successful completion of a cycle. These pop-up screens are shown below.

Table of Contents

Related product manuals