Maintenance Procedures Siderail
93
HEAD END SIDERAIL CABLE REPLACEMENT (CONTINUED)
6. Using side cutters, clip the cable ties (A) holding the cables together.
7. Using a #2 Phillips screwdriver, remove the cable clamp (C) from the siderail.
8. Disconnect cable (B) from the circuit board and cable (D) from the speaker.
NOTE
The speaker and nurse call are optional equipment and may not be in the siderail as shown.
9. Pull the cables through the siderail (toward the center of the bed).
10. Unplug the cable assembly (F) underneath the head section.
11. Reverse the above steps to install the new cable.
CAUTION
Be sure to position the cables on both sides of the pivot arm, as shown in the illustration on page 92, before
reattaching the pivot arm cover. If not done properly, the cover will not fit tightly and damage could occur to
the cables.