Marker Name Length in words. Field is optional and is absent for
MarkerType = 2.
Name of the marker. In case of odd number of MarkerName bytes
last byte is 0x00. Field is optional and is absent for MarkerType =
2.
length of the block together with IDs, [words]
B.2.11 Record with the state of the alarm markers
The record with the wave file name consists of six words:
<0xC5nn>
<AlarmMarkerStateP>
<AlarmMarkerState>
<AlarmMarkerSMS>
<AlarmMarkerEmail>
<0xCDnn>
in which:
nn - size of records,
Each of word <AlarmMarkerStateP>, <AlarmMarkerState>, <AlarmMarkerSMS> and
<AlarmMarkerEmail> denotes the state of the markers:
b15 = state of #16 alarm marker
...
b1 = state of #2 alarm marker
b0 = state of #1 alarm marker
<AlarmMarkerStateP> denotes the state of point markers
<AlarmMarkerState> denotes the state of continuous markers
<AlarmMarkerSMS> denotes the state of SMS markers
<AlarmMarkerEmail> denotes the state of Email markers
B.2.12 Record with the directivity results
length of the block together with IDs, [words]
eCompass result which is a deviation from North direction [-179°-
180°]. This value should be treated as a correction for DirectionXY
result
direction of the maximum energy noise source for X-Y axis [0-
359°]. -1 value means no dominant direction
energy calculated for DirectionXY to entire energy ratio [*10000]
direction of the maximum energy noise source for Z axis [0-180°].
-1 value means no dominant direction
energy calculated for DirectionZ to entire energy ratio [*10000]
energy distribution calculated for i-direction to entire energy ratio
for X-Y axis [*10000]