Automatic system check
When the Automatic system check feature is enabled the calibration factor of the instrument will be
periodically verified using built-in electrostatic actuator. You can specify the hour and a weekday of the
automatic system checks.
To activate the automatic system check, select Enabled (no Cal. Factor update) option from the
Mode dropdown list.
To schedule automatic system checks, choose weekday and select time of the system check from the
Do system check every day at (hh:mm) dropdown list.
Note: System check cannot be considered as a calibration. Calibration factor will not be
updated during Automatic System Check procedure.
Note: Measurements are not stopped for the Automatic System Check procedure.
Calibration logging
Results of calibration (both manual and automatic) and automatic system check are stored in the
calibration and system check history file on the instrument’s SD-card. Log file name is C.TXT. The file
can be downloaded to the PC and then opened in any text editor, such as MS Notepad. Maximum
allowed size of log file is specified in the Maximum calibration log file size [MB] position.
9.8.12 Alarm settings
In the Alarm configuration section, press Advanced Alarm Configuration button to configure alarm
functionality of SV 200A. In the Advanced Alarm Configuration window, you can define and configure
the set of alarm Conditions and Events.
Alarm Conditions and
Events tabs
Threshold conditions
configuration panel