9.8.9 Remote communication
The Remote communication tab enables configuring a remote communication with SV 200A via 3G
or WLAN/LAN. Also, some other additional functions can be configured in this tab: Bluetooth, GPS and
energy saving by means of switching off powering of modems or enabling powering during defined time
window.
Note: Use this tab carefully, because if you make improper changings and send new
settings to your SV 200A, your Internet connection may not work. And what is worse you
will not be able to change settings remotely!
The connection configuration can be also done through the Instrument wizard panel (see Chapter 9.3).
The Remote communication tab allows configuring all communication types in one place.
GSM Modem configuration
Connection mode defines the type of TCP/IP connection:
TCP server or TCP client. If it is None, there is no
connection with the Internet.
Typical GSM Modem configuration for connection with the
SvanNET web-service is presented right-hand.
Other configurations are dedicated to the specific user
needs or GSM operator conditions in some countries
(please refer your local GSM operator for details).
WLAN/LAN connection
configuration
3G connection configuration