Weighting filters (see C.3)
Z meeting requirements of the IEC 61672-1:2013 standard for the Class 1 Z“ filter
A meeting requirements of the IEC 651 and IEC 61672-1:2013 standard for the Class 1 “A” filter
C meeting requirements of the IEC 651 and IEC 61672-1:2013 standard for the Class 1 “Cfilter
B meeting requirements of the IEC 651 and IEC 61672-1:2013 standard for the Class 1 “Bfilter
Table C.1.2. Self-generated noise for different weighting filters
Electrical *)
Acoustical compensated
Weighting filter
A
C
Z
A
C
Z
Noise
< 13 dB
< 14 dB
< 15 dB
< 15 dB
< 15 dB
< 20 dB
*) measured with the ST 02 microphone equivalent impedance 18 pF +/-10%
Special filters (see Section C.1)
Frequency response of SV 200A is compensated by means of two digital filters:
Environmental compensation filter improving the complete instrument
frequency response in the free field for the reference
acoustic wave incidence angle 90 deg
Airport compensation filter improving the complete instrument
frequency response in the free field for the reference
acoustic wave incidence angle 0 deg
RMS detector
Digital “True RMS“ with Peak detection,
Resolution 0.1 dB
Range 327.7 dB
Crest Factor unlimited (for signals in 20 kHz band).
Overload detector
The instrument has the built-in overload detectors. The overload in the measurement channel (in its
analogue part) and the overload of the analogue / digital converter are both detected. The “overload”
indication is when the input signal amplitude is 0.5 dB above the declaredPeak measurement range”.
Underrange detector
The instrument has the built-in underrange detector. The “underrange” indication appears when the
minimum value of the RMS detector output goes below the specified lower linear operating range.
Time weighting characteristics (Exponential averaging)
Slow S“ according to IEC 61672-1:2013 Class 1, Equivalent Time Constant 1000 ms
Fast F” according to IEC 61672-1:2013 Class 1, Equivalent Time Constant 125 ms
Impulse I” according to IEC 60804:2000 Class 1, Equivalent Time Constant 35 ms, Hold Time
1500 s
Reference conditions
Class of the acoustic field Free field
Reference acoustic pressure 114.0 dB (related to 20 Pa)