EasyManuals Logo

Svantek SV 200A User Manual

Svantek SV 200A
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
37
SV 200A User Manual
Note: Unlike Calibration procedure, system check does not change the calibration factor of the
instrument.
The electrostatic actuator generates 1 kHz tone equivalent to sound pressure level of 94 dB (re. 20 μPa).
It can be turned on or off and programmed remotely by the user:
- via SvanNET Web service, in the Status tool (see Chapter 8.3) or
- via SVAN PC++ Remote Control software, in the Live Results window (see Chapter 9.6).
Auto System Check feature of SV 200A enables configuring and scheduling automatic check of the
instrument:
- via SvanNET Web service, in the Automatic system check panel (see Chapter 8.4.1) or
- via SvanPC++ Remote Control software, in the Automatic system check panel (see
Chapter 9.8.11).