Time history results are saved in a file with automatically defined name, which consists of a prefix (a
string of letters) and a number (string of digits). New time history is recorded in a new file, which name
is generated on the inherit principle – new file name has the same prefix as the previous file, but it’s
number is increasing by one.
Note: The Station(s) configuration window doesn’t allow changing names of the logger
files. But you can do it in the SVAN Files window with the help of the Setup file editor
function – see “SvanPC++ User Manual”.
Logger On
The Logger On position switches on/off the functionality,
which enables saving selected results from three user profiles
with the Logger step and Summary results with the step
specified by the Integration period parameter.
Note: When Logger On is deselected, no measurement results will be stored in the
SV 200A memory. Usually this setting should be on.
Logger step
The Logger step defines the period of logger results
integration/measurement and logging in the logger file. It can
be set from 2 milliseconds to 1 second in 1, 2, 5 sequences,
from 1 second to 59 second, from 1 minute to 59 minutes and
1 hour.
Logger meteo
The Logger meteo check-box controls saving results from
the SP 275 weather station (supplied as an option). The
results are saved with the Logger step period, unless it is
smaller than 1s. In such case, weather data will be stored in
the result file each second.