EasyManua.ls Logo

Tektronix DSA70000D Series - Set up a Custom Serial Bus; Set up a Pcie Serial Bus

Tektronix DSA70000D Series
1002 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
How to ? Set up a Custom serial bus
Set up a Custom serial bus
First, follow the common steps to set up any bus (see page 580). To continue to set up a custom serial
bus, follow these steps:
1. To dene the bus type in the B us Setup window, click Serial as the Bus Type, and select Custom
from the drop-down list.
2. To select the custom decoder, click the Custom Decoder arrow, and select the custom decoder from the
drop-down list.
3. To select the input for the displayed components of the bus, click on the component Input and select a
channel from the cascading list.
4. To set an input voltage reference threshold level, double-click the Threshold entry box and use the
keypad to enter the level.
5. Click the Display tab, and congure the bus display
(see page 590).
Set up a PCIe serial bus
First, follow the common steps to set up any bus (see page 580). To continue to set up a PCIe serial
bus, follow these steps:
1. To dene the bus type in the Bus Setup window, click Serial as the Bus Type, and select PCIe from the
drop-down list.
2. To select the data rate, click the Data Rate arrow, and select the rate from the drop-down list.
3. To select the input for the displayed components of the bus, click on the component Input and select a
channel from the cascading list.
4. To set an input voltage reference threshold level, double-click the Threshold entry box and use the
keypad to enter the level.
5. To set lane, double-click the Lane entry box and use the keypad to enter the lane.
6. Click the Display tab, and congure the bus display
(see page 590).
586 DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series