EasyManua.ls Logo

Tektronix DSA70000D Series - Trigger on a B Event

Tektronix DSA70000D Series
1002 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
How to ? Trigger on a B Event
Trigger on a B E vent
Use the following trigger sequence to trigger the instrument after a specied number of B events:
1. Set the A Trigger Type and Source from the A Event tab in the Trigger control window.
2. Set the B Trigger Type and Source from the B Event tab.
NOTE. For MSO70000C Series instruments, when you select a Pattern or a State trigger as the B E vent,
digital channels are not available.
3. Open the A->B Seq tab; then click Trig on nth Event.
4. Click in the Trig Event entry box; then enter the number of B events that should occur before starting
the acquisition after the A event.
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 643

Table of Contents

Related product manuals