EasyManua.ls Logo

Tektronix DSA70000D Series - Trigger on a Sequence

Tektronix DSA70000D Series
1002 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
How to ? Trigger on a sequence
Trigger on a sequence
Use the following procedure to set up the instrument to trigger on a sequence:
1. Connect a signal to an input c hannel.
2. Set the acquisition system to Run; then set the vertical and horizontal controls appropriately for the
signal you want to acquire.
3. To trigger on A Event (main) only:
a. Click Trigger on the toolbar; then open the A->B Seq tab of the Trigger control window.
b. Click the A Only button to turn off sequential triggering.
4. To trigger on a B event after time:
a. To set the time base to run after an A trigger, a trigger delay, and a B trigger, click Trigger on the
toolbar; then open the A->B Seq tab of the Trigger Setup window.
b. Click Trig After Time.
c. To set the trigger delay, click in the Trig Delay entry box; then use the multipurpose knob or
pop-up keypad to set the time.
d. To set the B trigger level, click in the B Trig
Level entry box; then use the multipurpose knob
or pop-up keypad to set the level.
5. To trigger on B events:
a. To set the time base to trigger after an A trigger and a specied number of B trigger events, click
Trigger on the toolbar; then open the A->B Seq tab of the Trigger Setup window.
b. Click Trig on nth Event.
c. To set the number of B trigger events, click in the Trig Event entry box; then use the multipurpose
knob or pop-up keyboard to set the number of events.
d. To set the B trigger level, click in the B Trig Level entry box; then use the multipurpose knob
or pop-up keypad to set the level.
6. To set up B Event triggering:
a. To set the B Event trigger, click Trigger on the toolbar; then open the B Event tab of the Trigger
control window.
b. To specify which channel becomes the B trigger source, click Source; then select the source
from the drop-down list.
c. Click Coupling, then select the trigger coupling to match your trigger signal.
d. To specify the direction of the edge, click a Slope button, positive or negative.
e. To set the B trigger level, click in the Level entry box; then use the multipurpose knob or pop-up
keypad to set the level.
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 641

Table of Contents

Related product manuals