RiverPro and RioPro ADCP Guide October 2017
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Page 103
EZ Sensor Source
Purpose Selects the source of environmental sensor data.
Format EZcdhprst
Default EZ1011101
Recommended Setting. The default setting for this command is recommended for most applications.
Range Firmware switches (see description)
Description Setting the EZ-command firmware switches tells the RiverPro/RioPro to use data from a
manual setting or from an associated sensor. When a switch value is non-zero, the
RiverPro/RioPro overrides the manual E-command setting and uses data from the appro-
priate sensor. If no sensor is available, the RiverPro/RioPro defaults to the manual E-
command setting. The following table shows how to interpret the sensor source switch
settings.
Table 15: Sensor Source Switch Settings
Field Value = 0 Value = 1
c Speed Of Sound Manual EC Calculate using ED, ES, and ET
d Depth Manual ED N/A
h Heading Manual EH Internal Heading Sensor
p Pitch (Tilt 1) Manual EP Internal Pitch Sensor
r Roll (Tilt 2) Manual ER Internal Roll Sensor
s Salinity Manual ES N/A
t Temperature Manual ET Internal Transducer Sensor
Example EZ1011101 means calculate speed of sound from readings, use pressure sensor, trans-
ducer heading, internal tilt sensors, and transducer temperature.