October 2017 RiverPro and RioPro ADCP Guide
Page 94
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
CW Output the Last Stored Ensemble
Purpose Requests the most recently stored ensemble for output.
Format CW
Recommended Setting. Use the WinRiver II software - Data Recovery.
Description Recalls the most recently stored ensemble for output. This command is only valid during
a deployment when recording is enabled and will give an error message until valid data is
available. The CW command is available for PD0 output formats, and follows the CF com-
mand setting for binary or Hex-ASCII output.
CZ Put the system to sleep
Purpose Tells the RiverPro/RioPro to power down.
Format CZ
Recommended Setting. This command should be used whenever batteries have been connected and
commands to start a deployment are not sent.
Description Sending the CZ-command powers down the RiverPro/RioPro. RiverPro/RioPro pro-
cessing is interrupted and the RiverPro/RioPro goes in the STANDBY mode (RAM is
maintained).
If the RiverPro/RioPro is running a sensor test under battery power, a fully charged battery
will be discharged in a few days.
>cz
Going to sleep