October 2017 RiverPro and RioPro ADCP Guide
Page 24
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
When connecting the RiverPro/RioPro I/O cable, do not apply any upward force on the
connector as it is being pulled off. Applying an upward angle as the cable is connected puts
stress on the end-cap connector. This may cause several serious problems:
1) The end-cap connector or connector pins can crack.
2) The O-ring on the bottom of the end-cap connector can be damaged.
3) The molded urethane on the end-cap connector may separate from the brass insert.
If the end-cap connector is damaged in any of these ways, the RiverPro/RioPro will flood.
Using the Cable Clips
If the retaining strap on the dummy plug or cable breaks, use the cable clips:
1. Remove the broken retaining strap.
2. Snap the clip onto the cable or dummy plug.
Black clip & 2-137 O-ring = cables
White clip & 2-130 O-ring = dummy plugs
3. Route the O-ring through the clip. Connect the
cable/dummy plug and then stretch the O-ring
over the connector.
75ZK6001-00 Kit, Clip, Cable
75ZK6001-01 Kit, Clip, Dummy Plug