October 2017 RiverPro and RioPro ADCP Guide
Page 120
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Timing Commands
The following commands set the timing of various profiling functions.
Available Timing Commands
This section lists the available Timing commands.
>T?
Available Commands:
TE 00:00:00.00 ---------- Time Between Ensembles
TF --/--/--,--:--:-- ---- Set First Ping Time (yy/mm/dd,hh:mm:ss)
TP 00:00.00 ------------- Time Between Pings
TS 14/09/30,09:56:51.62 - Set System Date and Time (yy/mm/dd,hh:mm:ss)
T? ---------------------- Display T-Command Menu
>
Timing Command Descriptions
TE – Time Per Ensemble
Purpose Sets the minimum interval between data collection cycles (data ensembles).
Format TEhh:mm:ss.ff
Range hh = 00 to 23 hours
mm = 00 to 59 minutes
ss = 00 to 59 seconds
ff = 00 to 99 hundredths of seconds
Default TE00:00:00.00
Recommended Setting. The default setting for this command is recommended for most applications.
Description During the ensemble interval set by TE, the RiverPro/RioPro collects one automatic en-
semble. If TE = 00:00:00.00, the RiverPro/RioPro starts collecting the next ensemble
immediately after processing the previous ensemble.
Example TE01:15:30.00 tells the RiverPro/RioPro to collect data ensembles every 1 hour, 15
minutes, 30 seconds.
The time tag for each ensemble is the time of the first ping of that ensemble.