Page 127
WO Number of SubPings
Purpose Controls the number of SubPings in Mode 12 water profiling.
Format WOx
Range x = 1 to 100 sub-pings
Default WO001
Recommended Setting. Special applications only.
Description: In Mode 12, a number of sub-pings are transmitted very rapidly and their results are aver-
aged internally to form a single Mode 12 ping.
This command has no effect unless the WM command is set to WM12.
WP Number of Pings
Purpose Sets the number of pings to average in each data ensemble.
Format WPnnn
Range nnn = 0 to 999 pings
Default WP001
Recommended Setting. Set using WinRiver.
Description WP sets the number of pings to average in each ensemble before sending/recording the
data.
1. If WP = zero the RiverPro/RioPro does not collect water-profile data.
2. The RiverPro/RioPro automatically extends the ensemble interval (TE) if WP x TP > TE.
WS Bin Size
Purpose Selects the volume of water for one measurement cell (bin).
Format WSnnn
Range nnn = 2 to 500 cm
Default WS005
Recommended Setting. Set using WinRiver.
Description The RiverPro/RioPro collects data over a variable number of bins. WS sets the size of
each bin in vertical centimeters.