Page 87
Bottom Track Commands
The RiverPro/RioPro uses these commands for bottom-tracking applications. Bottom track commands
tell the RiverPro/RioPro to collect speed-over-bottom data and detected range-to-bottom data.
Available Bottom Track Commands
This section lists the most often used Bottom Track commands.
>B?
Available Commands:
BP 1 -------------------- Number of BT Pings in ensemble [0-999]
BX 00500 ---------------- Max Depth (dm) [10-65535 dm]
B? ---------------------- Display B-Command Menu
Bottom Track Command Descriptions
BP Number of BT Pings
Purpose Sets the number of bottom-track pings to average together in each data ensemble.
Format BPnnn
Range nnn = 0 to 999 pings
Default BP001
Recommended Setting. The default setting for this command is recommended for most applications.
Description BP sets the number of bottom-track pings to average together in each ensemble before
sending/recording bottom-track data.
The RiverPro/RioPro interleaves bottom-track pings with water-track pings (see TP
command). If BP = zero, the RiverPro/RioPro does not collect bottom-track data.
The RiverPro/RioPro automatically extends the ensemble interval (TE) if BP x TP > TE.