RiverPro and RioPro ADCP Guide October 2017
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Page 45
Figure 16. RiverPro/RioPro Compass Calibration Screen Pitch/Roll
The Pitch/Roll calibration requires eight rotations while pitching the RiverPro/RioPro up and down.
For the bottom row of squares, pitch the RiverPro/RioPro between -22.5 to -45 degrees.
For the second row, pitch the RiverPro/RioPro between -22.5 to 0 degrees.
For the third row, pitch the RiverPro/RioPro between 0 to 22.5 degrees.
For the top row, pitch the RiverPro/RioPro between 22.5 to 45 degrees.
Each row can be completed during one rotation or you can vary the pitch as you rotate. A
good compass calibration requires slow, smooth movement to allow the compass to collect
data at each point.
Yearly Maintenance Items
The O-rings, desiccant and hardware should be replaced every one to two years.
Maintenance Kit
Table 2 and Table 3 lists the items in the maintenance kits. These kits are required when the RiverPro/Ri-
oPro has been opened. The maintenance kit includes the following tools and spare parts.
The maintenance kits are not included with the RiverPro/RioPro system. They are required to
properly close the system.
Order kit number 75BK6032-00 for RiverPro systems.
Order kit number 757K6078-00 for RioPro systems.
Table 2: RiverPro Maintenance Kit
Item # Part # Description
1. DES2 Desiccant
2. 5020 Silicone lubricant, 4-pack
3. 84Z-6000-00 Tool bag
4. 972-6001-00 O-ring, bore, 2-159
5. 972-8044-00 O-ring, face, 2-253
6. M4ALLENDRIVER M4 ball wrench