RiverPro and RioPro ADCP Guide October 2017
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Page 109
PC4 Display Voltage Monitor ADC Data
The PC4 Batt (V) value is ~ 0.8 volts below the supply voltage. VDD1 and VDD3 should be within 0.01
volts of the values shown in the example data.
>pc4
Battery ADC data is sampled and displayed in a loop.
Press any key to exit the loop.
Count Chan0 Chan1 Chan2 Chan3 Batt(V) Batt(A) VDD1 VDD3
\ 11 0685 00c6 0d49 0b7f 11.878 0.121 3.328 1.797
PC40 Display Scrolling Voltage Monitor ADC Data
PC40 is the same as PC4 except the data is displayed in a repeating loop at approximately 0.5-sec update
rate.
>pc40
Battery ADC data is sampled and displayed in a loop.
Press any key to exit the loop.
Count Chan0 Chan1 Chan2 Chan3 Batt(V) Batt(A) VDD1 VDD3
0 0684 00ca 0d49 0b80 11.871 0.123 3.328 1.797
0 0683 00e2 0d49 0b7f 11.865 0.138 3.328 1.797
0 0684 00ba 0d49 0b80 11.871 0.114 3.328 1.797
0 0684 00b2 0d49 0b7f 11.871 0.109 3.328 1.797
>
PC5 Display BlueTooth RSSI Data
TRDI use only.
PC50 Display Scrolling BlueTooth RSSI Data
TRDI use only.
PC6 Display GPS RSSI Data
TRDI use only.
PC60 Display Scrolling GPS RSSI Data
TRDI use only.