Specifications
Table 2 Specifications for Spectronic GENESYS 6 UV-Visible spectrophotometer
Feature Specifications
Spectral bandwidth 1.8nm
Optical System Split-beam, grating-based, dual detectors
Lamp source;
Lifetime
Xenon;
5 years typical
Wavelength:
Range 190 - 1100nm
Accuracy ± 1.0nm
Repeatability ± 0.5nm
Display 320 x 240 pixel backlit LCD, 3.8” x 2.8”
Photometric:
Range 0.3 - 125%T; -0.1 - 3.0 A; ± 9999C
Readout Absorbance, Transmittance, Concentration
Accuracy
1
0.5% or 0.005A, whichever is greater, up to 2A
Noise
≤ 1mA at 0A; <– 2mA at 2A, peak-to-peak at 340nm
Drift
≤ 1mA/hour
Stray light
2
≤ 0.1%T at 220, 340 and 400nm
Standard Interface Bi-directional RS232C; parallel output in HP PCL format
Cell Holder Automatic 6-cell positioner
Keypad Membrane keypad
Scanning Speed 10 -- 2,000nm/min
Slew rate 11,000nm/min
Scanning Sampling Interval 0.1, 0.2, 0.5, 1.0, 2.0, 5.0nm
Disc Drive 3.5" floppy; 1.44 Mb
Test Storage Up to 80 tests in non-volatile memory; up to 256 on a floppy disk
Languages (user selectable) Software and Printout: English, French, German, Spanish, Italian
Internal Printer (optional) 40-column graphical
Data Storage Floppy disk
Operator’s manual: English, French, German, Spanish, Italian
Power Requirements Selected automatically; 100 - 240 Volts
Dimensions 330W x 410D x 235H mm (13" x 16" x 9")
Weight 9.1 Kg (20.1 lb)
Warranty 1 year
1
Measured using NIST 930D filters
2
220nm, 340nm and 400nm measurements with SPECTRONIC Standards (#333150)