Technical Support Bulletins
STANDBY x
Set the number of minutes for the standby function. 0 is used to disable the
standby function:
x may only be one of the following discreet values:
0, 15, 30, 60, 90
TIME
Report the current time and date
TIME x
Set the current time and date:
hhmm MMDDYY W
hh hours 0 <= hh <= 23
mm minutes 0 <= mm <= 59
MM month 1 <= MM <= 12
DD date 1 <= DD <= 31
YY year 0 <= YY <= 99
W day 1 <= W <= 7 1 is Sunday
VER
Report the instrument’s software version:
1.000
Immediate Commands: Special control-keys that are executed immediately
^C
This command will reset the instruments serial port. All characters in both the
input and output queues will be lost
^S
Upon receipt of this character all output from the instrument will be suspended
^Q
Upon receipt of this character serial output from the instrument will resume
^R
Reset the instrument. The same as a power-on reset, except hardware that
depends on the hardware RESET line may not be restarted