EasyManuals Logo

Thermo Spectronic BioMate 3 User Manual

Default Icon
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
Optical Alignment
To install and adjust any of the mirrors:
1. Make sure the proper mirror has been selected for the location and is oriented with the mirrored
surface facing in the correct direction as shown in Figure 6.2. (All mirrors face in toward the center
of the instrument.)
2. Insert the two mounting screws through the foot of the mirror into the base. Snug the mirrors
slightly but do not tighten. The mirror should be able to be rotated in the hole with a slight
resistance. The screws must be loose enough so that the mirror will slide under the screw. Never
force or twist the mount to align it.
3. Rotate the mirror as required until it is properly aligned horizontally.
4. Tighten the two mounting screws.
5. Tighten or loosen the tilt adjustment screw as necessary until the vertical alignment of the mirror is
proper.
Alignment Sequence
The various components of the optical path and their locations are shown in Figure 6.2. In addition to any
replacement components, which may be required, the following procedure requires the use of the
Spectronic GENESYS 10 Service Alignment Tool Kit [335901-400], which includes the following
components:
335901-401 Mirror B2 Target
335901-408 Dental Mirror
335901-403 zero order grating Tool
335901-406 Spectronic GENESYS 10 Sample Target
335901-407 Spectronic GENESYS 10 Mirror Adjust Tool
H-152 728 Spacing Tool
Figure 6.2 Optical Path Components
6-3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Spectronic BioMate 3 and is the answer not in the manual?

Thermo Spectronic BioMate 3 Specifications

General IconGeneral
BrandThermo
ModelSpectronic BioMate 3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals