Technical Support Bulletins
Cable Replacement Instructions
Spectronic™ GENESYS™ 10
(Cat. Nos. 335900-335902)
Spectronic BioMate™ 3
(Cat. Nos. 335904-335905)
Warning
Follow instructions below very carefully. Failure to do so may result in instrument damage and/or
personal injury.
Disconnect power from the instrument before servicing.
Observe proper ESD protocol.
When installing the following in a Spectronic GENESYS 10 or Spectronic BioMate 3:
335901-657 Hantronix display (
13-pin header): ensure that you use the 335901-1608 display cable
(Black wires,13-pin connector at P20 end).
335901-657 EDT or Hantronix display (
14-pin header): ensure that you use the 335901-1612 display
cable (White wires, 14-pin connector at P20 end).
Ground wires and the printer cable assembly are also included in this kit; no special instructions
apply.
11-6