Figure 100. Results
from the Reset Zero
Offset procedure are
displayed on this
popup.
Figure 101. The Reset
Zero Offset option
should be executed
prior to performing
the Zero Calibration
process.
DIAGNOSTICS
FMS-1650
TRIATEK reserves the right to change product specifications without notice.
- 54 -
Zero Calibrating the FMS-1650