132
13 Options
132
Options
Chapter
13
■ Continuous transmitting format
- Transmitting format 1
- Transmitting format 2
*: When analog peak hold (peak hold selection = 1: analog, and hold mode = 1: peak hold) is used,
the 5 digits are the same as the linked indicated values.
- Transmitting format 3
*1 O: Overload
(LOAD, OFL, OVERLOAD)
S: Stable
M: Unstable
H: Hold
Priority: H > O > (S or M)
*4 N: Near zero OFF
Z: Near zero ON
*5 H: HH limit ON
L: LL limit ON
G: HH limit and LL limit OFF
N: HH limit and LL limit ON
F: Compare OFF
Priority: N > (H or L) F > G
*2 A: Zero tracking OFF
T: Zero tracking ON
*3 H: HI limit ON
L: LO limit ON
G: HI limit and LO limit OFF
N: HI limit and LO limit ON
F: Compare OFF
Priority: N > (H or L) F > G
*6 Decimal place
3: 0.000
2: 00.00
1: 000.0
0: None
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
GS ,*1 *2 *3 *4,, ,,
Indicated value linked (5 digits + decimal point)
Sign
+ or -
±
*5 ,
20 21
Terminator
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
W T *1 *2 *3 *4 ±
Indicated value not
Sign
+ or -
Indicated value
*5
SOH STX ETX BCC
Sign
+ or -
±
*6
22 23
BCC calculation range
Each ASCII code of SOH, STX, ETX
SOH: 01
STX: 02
ETX: 03
BCC calculation method
For BCC,
each code in the BCC calculation range is represented in
hexadecimals and is XORed with all data.
linked (5 digits) linked (5 digits)*
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
GS ,*1 *2 *3 *4,, ,
Indicated value linked (5 digits + decimal point)
Sign
+ or -
±
,
Ter min at or