61
7 Settings and Operations Related to Hold
61
Settings and Operations Related to Hold
Chapter
7
●Section definition peak hold
7-5. Hold detection wait
Detection wait of hold input is set. Changes of hold input will not be recognized within the time set
as detection wait after capturing the changes (ON -> OFF, OFF -> ON). This function is useful to
ignore chattering parts.
■Hold detection wait setting method
1.
Select setting mode 5.
→ (four times)
2. Select "hold detection wait".
Press the key eight times.
Set the hold detection wait using the
(numerical value input) and (shift) keys, and
confirm it with the key.
3. Press the key to return to the indicated value display.
㪘㫋㩷㪻㪼㫋㪼㪺㫋㫀㫆㫅㩷㫊㫋㪸㫉㫋
㪘㫋㩷㪻㪼㫋㪼㪺㫋㫀㫆㫅㩷㫊㫋㫆㫇
HOLD
OFF
ON
H.RESET
OFF
ON
Hold
OFF
ON
HOLD
OFF
ON
H.RESET
OFF
ON
Hold
OFF
ON
Hold values are canceled at
of numerical values during
detection start; therefore, changes
Hold values are maintained during
of hold values are shown.
detection; therefore, only changes
Once H.RESET is entered, current
value display returns.
detection are shown.
Hold detection wait
(0.00 to 1.00)