9.3 Configuring filtering factor
Figure 60 Filtering factor configuration in Insight PC software
1 Enable Filtering on/o to use the default filtering factor.
2 Enable Extended filtering on/o if you want to enter a custom filtering factor value.
3 After enabling the Extended filtering on/o selection, enter the custom value into the
Filtering factor field.
You can apply a filtering factor to the measurement output to average out measurement
output over time (for more information, see Environmental compensation and measurement
settings overview (page 114)). A default or custom filtering factor can be set using either the
local display interface or Vaisala Insight PC software.
To set a filtering factor:
1. If you are using the local display interface:
a. Select Menu > Settings > Measurement settings > Filtering.
b. To enable the default filtering factor (13 second moving average (T90)), set the
Filtering on/o selection to ON.
c. If you want to enter a custom filtering factor, set the Extended filtering on/o
selection to ON, select Filtering factor and enter the custom value with the Up and
Down buttons.
d. After the final digit, pressing Select saves the pressure value.
HMT370EX User Guide M212305EN-B
116