10. Calibration and adjustment
10.1 Calibration and adjustment overview
Figure 61 Insight PC software calibration view example
Figure 62 Local display calibration menu
The RH and T measurement of the probe used in the HMT370EX transmitter can be calibrated
and adjusted (1-point or 2-point adjustments) using one of the following alternative
configuration options:
Detaching the probe body from the transmitter and connecting the probe to Insight PC
software (requires M12 Indigo USB Adapter cable accessory (Vaisala item USB2)).
Calibrating and adjusting the probe with Insight PC software when the probe is attached
to the transmitter (requires M8 USB cable accessory 219690).
Calibrating and adjusting the probe with the local display interface of the transmitter
when the probe is attached to the transmitter.
Chapter 10 – Calibration and adjustment
119